This is a working overview of Chromatogram, written for readers who want more than a one-paragraph summary but less than a textbook.
This page was last updated on 2026-06-22 and is reviewed periodically as new material appears.
High-performance liquid chromatography is an analytical technique that separates components in a liquid sample. A pump moves a liquid mobile phase through a column packed with a solid stationary phase. Compounds interact differently with both phases and travel at different rates, leaving the column at distinct retention times. A detector records these arrivals as peaks on a chromatogram. The resulting pattern supports identification and quantification of substances in mixtures. Modern instruments use high pressure to force solvent through small particles, which improves speed and resolution compared with older low-pressure liquid chromatography methods.
Separation in HPLC depends on the chemistry of the stationary phase, the composition of the mobile phase, and the physical properties of the column. Reverse-phase separations use a nonpolar stationary phase and a polar mobile phase, and they are common for many organic compounds. Ion-exchange, size-exclusion, and normal-phase modes serve other classes of analytes. Gradient elution changes solvent strength over time, while isocratic elution holds it constant. Flow rate, temperature, particle size, and column length all influence peak shape and resolution. Detection may use ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry, depending on the analyte and the required sensitivity.
Routine HPLC testing compares a sample result with a calibration curve prepared from known reference standards. Peak area or peak height is plotted against concentration, and the curve is used to estimate unknown amounts. Retention time supports tentative identification when compared with a standard, though mass spectrometry or another confirmatory method may be needed for definitive identification. Pre-run checks verify repeatability, resolution, and peak symmetry before sample analysis. Limits of detection and quantification describe the smallest amounts that can be reliably observed or measured. Sample preparation, filtration, and degassing help prevent column damage and inconsistent results.
Instrumentation includes a solvent delivery system, an autosampler, a column oven, and one or more detectors. Reversed-phase columns with chemically modified silica are widely used, but normal-phase, ion-exchange, size-exclusion, and affinity modes exist for specific separations. Detectors may rely on ultraviolet absorbance, fluorescence, refractive index, or mass spectrometry. Column temperature, mobile phase composition, and flow rate are adjusted to improve resolution. System pressure is monitored because rising pressure can indicate column blockage or deteriorating packing.
Separation performance depends on particle size, pore size, column length, and the chemistry of the stationary phase. Smaller particles generally improve efficiency but require higher pressure and suitable instrumentation. The mobile phase often contains buffers and organic solvents that influence retention and selectivity. Testing labs select conditions based on the analytes, sample matrix, and required sensitivity. Method development frequently involves screening several columns and solvent mixtures before a final set of conditions is chosen.
| Property | Value | Notes |
|---|---|---|
| Common abbreviation | HPLC | High-performance liquid chromatography |
| Separation basis | Differential partitioning | Between liquid mobile phase and solid stationary phase |
| Common mode | Reverse phase | Nonpolar column, polar mobile phase |
| Typical detector | UV-Vis absorbance | Widely used for compounds with chromophores |
| Typical column particle size | 2–5 µm | Smaller particles can improve resolution |
Practical HPLC testing depends on careful sample preparation and instrument maintenance. Samples may require filtration, dilution, pH adjustment, or extraction to avoid column damage and matrix interference. Mobile phases are degassed and filtered, and columns are equilibrated before injection. Common problems include peak tailing, baseline drift, ghost peaks, carryover, and co-elution of analytes. Documentation of instrument logs, calibration records, and electronic audit trails supports data integrity and traceability. Ongoing training and routine maintenance help reduce variability between analysts and laboratories.
Quality control laboratories use HPLC to check identity, purity, concentration, and stability of raw materials and finished products. A validated method specifies the column, mobile phase, flow rate, detection wavelength, injection volume, and run time. Samples are prepared and compared against reference standards of known concentration. The resulting chromatogram provides quantitative data, such as assay values and impurity levels. This approach is common in pharmaceutical, food, environmental, and industrial testing where consistent measurements are required.
Method validation demonstrates that an analytical procedure is suitable for its intended purpose. Typical validation characteristics include accuracy, precision, specificity, linearity, range, limit of detection, limit of quantitation, and robustness. Regulatory guidance from bodies such as the International Council for Harmonisation and the United States Pharmacopeia outlines expectations, though specific criteria depend on the product and method. System suitability tests are run before sample analysis to confirm resolution, peak symmetry, column efficiency, and injection repeatability. Failure of these checks can invalidate a batch of measurements.
Routine HPLC testing depends on controlled reagents, calibrated instruments, and documented procedures. Columns degrade over time, so retention times and peak shapes are monitored for drift. Mobile phases are filtered and degassed to prevent pump damage and detector noise. Reference standards must be traceable and stored under suitable conditions. Data handling systems record injections, calculations, and audit trails. Quality control samples interspersed with unknowns help detect errors during a run.
Developing an HPLC method begins with defining the purpose, such as quantifying a main component, measuring impurities, or confirming identity. Analysts select separation mode, column, mobile phase, detection, and sample preparation based on analyte properties and matrix. Experiments vary solvent strength, pH, buffer type, and temperature to achieve resolution between critical peaks. The goal is a robust method that produces reliable results across instruments and operators. Method development often involves trial runs and statistical optimization.
Routine quality control monitors retention time shifts, baseline noise, system pressure, and peak shape. Trends can reveal column aging, mobile phase preparation errors, detector drift, or sample degradation. Corrective actions may include replacing the column, preparing fresh mobile phase, or recalibrating the detector. Stability testing often uses HPLC to measure parent compound loss and degradation product formation. Open questions remain about how accelerated stability results extrapolate to long-term storage under varied conditions.
Quality control for HPLC testing combines scheduled checks, documented procedures, and review of results. Before sample analysis, system suitability testing confirms that the instrument, column, and method meet predefined criteria. Common criteria include resolution between critical peaks, retention time precision, peak tailing, and theoretical plate count. Failure triggers investigation before results are reported. Records link raw data, calculations, instrument logs, and analyst identity to each batch, supporting audits and repeat analysis.
Method validation evaluates accuracy, precision, specificity, linearity, range, detection limit, quantitation limit, and robustness. Regulatory guidance for pharmaceuticals, foods, and environmental testing defines expected documentation and acceptance criteria. Verification confirms that a validated method works in a specific laboratory with its own instruments and reagents. Calibration curves use reference standards with known purity and traceability, while measurement uncertainty is estimated from validation data, control charts, and collaborative studies. The scope of validation depends on the method's intended use.
=== Amplification === Most mechanisms of symmetry breaking focus on amplification of an initial stochastic enantiomeric excess. The most likely path for this amplification step is by asymmetric autocatalysis. An autocatalytic chemical reaction is that in which the reaction product is itself a reactive, in other words, a chemical reaction is autocatalytic if the reaction product is itself the catalyst of the reaction. In asymmetric autocatalysis, the catalyst is a chiral molecule, which means that a chiral molecule is catalyzing its own production. An initial enantiomeric excess, such as can be produced by polarized light, then allows the more abundant enantiomer to outcompete the other.
Boltovsky, T., Bernardo Abiahy, Viviana A. Alder, Martin V. Angel, Renate Bernstein, Dennis Binet, Demetrio Boltovsky, Jean Bouillon, Janet Bradford-Grieve, John-Paul Casanova, Paul Cornelius, Jose R. Dadon, Christina Deponte, Graciela B. Esnal, Maria Alamo, Adilson Fransozo, Mark Gibbons, Ray Gibson, and Cristoph Helemben. "Marine Species Identification Portal : Catostylus tagi." Marine Species Identification Portal : Catostylus tagi. UNESCO, n.d. Web. 24 Sept. 2014.
== Dietary sources == Good sources of phenylalanine are eggs, chicken, liver, beef, milk, and soybeans. Another common source of phenylalanine is anything sweetened with the artificial sweetener aspartame, such as diet drinks, diet foods and medication; the metabolism of aspartame produces phenylalanine as one of the compound's metabolites.
Chikungunya vaccines are vaccines intended to provide acquired immunity against the chikungunya virus. As of 2025, two vaccines have been licensed in some countries. These are Ixchiq, a live attenuated vaccine from Valneva, and Vimkunya manufactured by Bavarian Nordic which utilises virus-like particle technology. The most commonly reported side effects of Ixchiq include tenderness at the injection site, as well as headache, fatigue, muscle pain, joint pain, fever, and nausea. However the license for Ixchiq has been suspended or restricted in some countries due to the risk of severe side effects, particularly in older people.
Sources: en.wikipedia.org
== Personal life == John Knox celebrated 60 years of marriage to his wife Josephine in March 2017. He died on 15 October 2018 in Edinburgh, aged 90. He is survived by his wife, four sons, 12 grandchildren, and 4 great-grandchildren.
Richmond – This rivalry stems out of the 1942 Grand Final which Essendon won. In 1974, a half-time brawl took place involving trainers, officials and players at Windy Hill and has become infamous as one of the biggest ever. The teams didn't meet in the finals between 1944 and 1995, but there have been many close margins in home and away season matches as a result of each team's "never say die" attitude and ability to come back from significant margins in the dying stages of matches. Having met in the AFL's Rivalry Round in (2006 and 2009) and meeting in the Dreamtime at the 'G match since 2005, the rivalry and passion between the clubs and supporters has re-ignited. In recent years the rivalry has been promoted as the "Clash of the Sash". Hawthorn – The two sides had a number of physical encounters in the mid-1980s when they were the top two sides of the competition. The rivalry was exacerbated when Dermott Brereton ran through Essendon's three-quarter time huddle during a match in 1988 and again by an all in brawl during a match in 2004 allegedly instigated by Brereton (now known as the Line in the Sand Match after the direction allegedly given by Brereton for the Hawthorn players to make a physical stand). This was reminiscent of the 1980s when battles with Hawthorn were often hard and uncompromising affairs. During Round 22 of the 2009 season, Essendon and Hawthorn played for the last finals spot up for grabs. The teams played out an extremely physical game and despite being 22 points down at half time Essendon went on to win by 17 points.
=== Stage II: Edema === This stage begins the acute phase of noma. The telltale sign is facial edema (swelling) of the lips, cheeks, eyes, etc. Ulceration of the gums worsens during this stage; ulceration may spread to the mucosa (soft, mucus-producing tissue) of the mouth and nose. The patient may feel pain or soreness in their mouth and cheeks. Other symptoms at this stage include fever, drooling, fetid breath, lymphadenopathy (swollen lymph nodes), and difficulty eating. Progression of the disease can be halted with appropriate treatment.
== Applications == Food-packaging applications include oxygen removal, moisture control, ethylene scavenging, antimicrobial activity, modified-atmosphere maintenance, freshness indication, and temperature monitoring. Pharmaceutical and medical applications include moisture-control closures, oxygen absorbers, temperature indicators, authentication features, adherence monitoring, and track-and-trace systems. Industrial applications include corrosion-inhibiting packaging for machinery, electronics, tools, automotive components, and military equipment. Shock and vibration monitoring is used with calibrated or fragile products. Agricultural applications include ethylene-control materials, humidity management, antimicrobial pads, freshness indicators, and controlled-release substances used during produce storage and distribution. Active packaging can also be used in transport systems for biological materials and organs, although these applications require specialised temperature control, sterility, and regulatory oversight.
Polycrystalline graphene was generated from these nucleation sites and was subsequently annealed at 3000K, and then quenched. Based on this model, they found that cracks are initiated at grain-boundary junctions, but the grain size does not significantly affect the strength. Second, in 2013, Z. Song et al. used MD simulations to study the mechanical properties of polycrystalline graphene with uniform-sized hexagon-shaped grains. The hexagon grains were oriented in various lattice directions and the GBs consisted of only heptagon, pentagon, and hexagonal carbon rings. The motivation behind such a model was that similar systems had been experimentally observed in graphene flakes grown on the surface of liquid copper. While they also noted that crack is typically initiated at the triple junctions, they found that as the grain size decreases, the yield strength of graphene increases. Based on this finding, they proposed that polycrystalline follows pseudo Hall-Petch relationship. Third, in 2013, Z. D. Sha et al. studied the effect of grain size on the properties of polycrystalline graphene, by modeling the grain patches using Voronoi construction. The GBs in this model consisted of heptagons, pentagons, and hexagons, as well as squares, octagons, and vacancies. Through MD simulation, contrary to the aforementioned study, they found an inverse Hall-Petch relationship, where the strength of graphene increases as the grain size increases. Experimental observations and other theoretical predictions also gave differing conclusions, similar to the three given above.
Sources: en.wikipedia.org
HPLC testing measures the presence and amount of one or more compounds in a liquid sample. It separates mixture components and records detector responses as peaks, which are compared with reference standards. Results are usually reported as concentrations or relative percentages.
Retention time is the interval between sample injection and the detector response for a given compound. It depends on the compound's interactions with the stationary and mobile phases under set conditions. Matching a retention time to a standard supports tentative identification but is not always unique.
HPLC alone can separate unknown compounds and provide retention times, but it often cannot identify them with certainty. Coupling HPLC to mass spectrometry gives mass information that improves identification. Confirmation usually requires comparison with reference standards or complementary techniques.
HPLC separates and detects individual compounds in a liquid sample, producing peaks at characteristic retention times. Peak area or height can be used to estimate concentration when calibrated with known standards. It does not identify unknown compounds with certainty unless additional detectors or reference materials are used.